Testing standard

ISO 7500-1

Metallic materials — Calibration and verification of static uniaxial testing machines — Part 1: Tension/compression testing machines — Calibration and verification of the force-measuring system

Written and technically reviewed by Dak System Inc. engineeringLast reviewed

ISO 7500-1 covers the calibration and verification of the force-measuring system on tension and compression testing machines. It assigns a class — 0,5, 1, 2 or 3 — where the class number is the maximum permissible relative accuracy error in per cent. The class belongs to a force range, not to the machine as a whole.

At a glance

Test type
Calibration & verification
Published by
ISO
Edition
ISO 7500-1:2018

What the test does

Force-proving instruments traceable through ISO 376 are installed in the machine's own load train, aligned as a specimen would be, and force is applied slowly through each range while the indicated force is compared against the true force. The series is repeated to establish repeatability, and zero readings and the resolution of the indicating system are recorded. Alongside this calibration sits a general inspection of the machine, including the accessories used to apply force. Each force range is then assigned a class — 0,5, 1, 2 or 3.

What it measures, and why it matters

Whether the number on the screen is the force in the specimen. Almost every test method on this site names ISO 7500-1 Class 1 as its accuracy requirement, and this is the standard behind that phrase: the class number is the maximum permissible relative accuracy error in per cent, so Class 1 means ±1 % of the indicated force. Accuracy alone does not earn the class — repeatability, zero error and resolution each carry their own limit, and a machine can meet the accuracy figure and fail on repeatability. The error is expressed as a percentage of reading, not of full scale, which is why holding a class at a tenth of capacity is much harder than at full capacity.

The classes

The class number is the permitted error in per cent. That is the whole scheme, and it is why a Class 1 machine is a machine good to one per cent.

Class 0,5
±0,5 % maximum permissible relative accuracy error
Class 1
±1 %What the great majority of test methods call for, which is why almost every page on this site names it.
Class 2
±2 %
Class 3
±3 %
Accuracy alone is not enough
Repeatability, zero error and resolution must also be within that class's limitsA machine can meet the accuracy figure and still fail the class on repeatability.
Assigned per force range, not per machine
This is the most misread line in the standard. A 100 kN frame can be Class 1 from 20 kN up and unclassified below it.

There is no such thing as a Class 1 machine in the abstract. There is a Class 1 range on a machine, with a lower limit, and testing below that limit is testing outside the certificate.

Test speed

Force application
Slow enough to avoid rate effects in the machine or the proving instrument
Reference
Force-proving instruments to ISO 376
Verification
A general inspection of the machine, including the force-application accessories, as well as the calibrationGrips, platens and couplings are in scope — the standard does not treat the load cell in isolation.
Static only
The calibration values are not necessarily valid for high-speed or dynamic testingStated in the standard. A fatigue frame certified to this has been certified standing still.

Calculations

Relative accuracy errorq

q = (Fi − F) / F × 100

Fi
force indicated by the machine
F
true force from the proving instrument

Expressed as a percentage of the true force, not of full scale. That distinction is why a percentage-of-reading specification gets tighter in absolute terms as the force falls, and why low-force work is hard.

Class

The class number is the maximum permissible relative accuracy error, in per cent

Class 1 permits ±1 %, Class 0,5 permits ±0,5 %. Repeatability, zero error and resolution each carry their own limit for the same class.

What a class does not tell you

The range it applies over

Always read the certificate's lower limit alongside the class. The two are meaningless apart.

How the test runs

  1. 01Establish which force ranges and which load cells are to be covered.
  2. 02Obtain force-proving instruments to ISO 376 covering those ranges.
  3. 03Inspect the machine generally, including the grips, platens and couplings used to apply force.
  4. 04Install the proving instrument in the load train, aligned as a specimen would be.
  5. 05Apply force slowly through the range, recording indicated against true force.
  6. 06Repeat the series as the standard requires, to establish repeatability.
  7. 07Record zero readings and the resolution of the indicating system.
  8. 08Calculate the relative accuracy error at each point.
  9. 09Check accuracy, repeatability, zero error and resolution against the limits for the intended class.
  10. 10Assign the class to each force range, with its lower limit.
  11. 11Issue the certificate stating the class, the range it covers, and the date.

The fixture this method needs

Self-identifying

Load Cells

The force-measuring system is what this standard classifies. A frame carries several interchangeable cells, and each one is calibrated and classified over its own range rather than inheriting the frame's certificate.

Specifications

What the report has to contain

  • Reference to ISO 7500-1 and the edition
  • Machine identification, capacity and the load cell verified
  • Identification and traceability of the force-proving instruments used
  • Ambient temperature
  • Indicated and true force at every point
  • Relative accuracy error, repeatability, zero error and resolution
  • The class assigned
  • The force range over which that class applies, including its lower limit
  • Whether the general inspection of the machine and its accessories passed
  • Date of calibration and the interval to the next

What the machine must be capable of

Being measured honestly. In practical terms that means a load train that accepts a proving instrument in the same alignment a specimen would occupy, an indicating system with enough resolution that the class is not resolution-limited, and grips, platens and couplings in good enough condition to pass the general inspection — they are explicitly in scope, and a worn coupling puts bending into the load train that no amount of load cell accuracy corrects. Where a frame carries several interchangeable cells, each is calibrated over its own range rather than inheriting the frame's certificate.

What goes wrong in practice

Quoting a class without its range, which is the single commonest error and makes the certificate unreadable. Testing at forces below the classified lower limit — easy to do when a large frame is used for a small specimen, and invisible unless someone checks. Treating a static certificate as covering dynamic work: the standard states that its values are not necessarily valid for high-speed or dynamic testing, so a fatigue frame certified under it was certified standing still. And assuming a good force certificate implies good results, when a modulus also depends on an extensometer certified under a different standard entirely.

ISO 7500-1 or ASTM E4

ISO 7500-1ASTM E4
StructureClasses 0,5 / 1 / 2 / 3A single ±1 % baseline, no class ladder
Reference instrumentsTo ISO 376Traceable to SI through a national metrology institute
Scope of verificationMachine inspection plus force calibrationForce calibration and verification
ResultA class, per rangeVerified, or not

Both give traceable force. ISO tells you how good the machine is; ASTM tells you whether it is good enough. Test methods name one or the other, and a laboratory serving both usually holds both certificates.

Questions we are asked about this test

What is ISO 7500-1?

It is the ISO standard for calibrating and verifying the force-measuring system of tension and compression testing machines. It covers a general inspection of the machine, including the accessories used to apply force, together with a calibration against force-proving instruments, and it assigns an accuracy class. The current edition is ISO 7500-1:2018, the fifth.

What does Class 1 actually mean?

That the maximum permissible relative accuracy error is ±1 % of the indicated force. The class number is the tolerance in per cent, which makes the scheme unusually easy to read: Class 0,5 is half a per cent, Class 2 is two. But accuracy alone does not earn the class — repeatability, zero error and resolution must each also fall within that class's own limits, so a machine can hit the accuracy figure and still fail.

Is my machine Class 1?

Strictly, no machine is. The class is assigned to a **force range**, not to a machine, and this is the most commonly misread thing in the standard. A 100 kN frame might be Class 1 from 2 kN to 100 kN and carry no classification below 2 kN. Testing a 500 N specimen on it is then testing outside the certificate, however impressive the certificate looks. Always read the class together with the lower limit of the range it covers.

Why is the error a percentage of reading rather than of full scale?

Because that is what the standard specifies, and it has a consequence worth understanding. A percentage-of-reading tolerance stays proportionally constant but shrinks in absolute terms as the force falls, so holding ±1 % at a tenth of capacity is far harder than at full capacity. It is precisely why low-force work — catheters, closures, fine films — needs a load cell sized for the force rather than for the frame.

Does this cover a fatigue machine?

Not while it is running. The standard says its calibration values are not necessarily valid for high-speed or dynamic testing. A frame certified under it has been certified standing still, so a dynamic test may deliver a force meaningfully different from the one commanded. That is why dynamic force verification at the test frequency is a separate exercise, and why fatigue standards ask for it.

Does it cover the grips as well as the load cell?

Yes, in the verification part. The standard requires a general inspection of the testing machine including its accessories for force application, so the grips, platens and couplings are in scope rather than the load cell being treated in isolation. A worn grip or a sloppy coupling can put bending into the load train that no amount of load cell accuracy will fix.

What about the extensometer?

Different standard. ISO 9513 covers the calibration of extensometer systems used in uniaxial testing. Force and extension are certified separately, which matters because a modulus depends on both — a machine with an immaculate force certificate and an uncalibrated extensometer will still produce a wrong modulus.

Running ISO 7500-1 on the Series 7200 and Series 9000

Dak verifies against whichever standard the method names, and where a class applies our frames sit a class tighter than it asks.

The method asks forDak supplies
CapacityWhatever the machine's ranges are — the point of the exercise is to classify each of themLoad cells from 1 kg to 60 ton on the Series 7200, and 0.5 to 100 kN on the Series 9000
Force accuracyForce-proving instruments to ISO 376ISO 7500-1 Class 0.5, verified to ASTM E4, DIN 51221 and BS 1610
GrippingForce-proving instruments to ISO 376, applied through the machine's own load trainWedge, vice-action, pneumatic and hydraulic grips, built to the specimen
Environment23 ± 2 °C standard laboratory atmosphere3009 series chambers, −150 °C to +400 °C — temperature only

This page describes the method as practised. The governing text is the current edition from the issuing body. Tell us what you are testing and we will answer with the machine, the fixture and a quotation.