Testing standard

ISO 7206-4 Endurance Testing of Stemmed Femoral Hip Components

Implants for surgery — Partial and total hip joint prostheses — Part 4: Determination of endurance properties and performance of stemmed femoral components

Written and technically reviewed by Dak System Inc. engineeringLast reviewed

ISO 7206-4 determines the endurance properties and performance of stemmed femoral components of hip joint prostheses. The stem is embedded to a defined level and loaded sinusoidally through the head at 1 Hz to 30 Hz, simulating the dynamic loading of gait. The standard fixes the requirement itself: 5 000 000 cycles at an endurance limit force of 2 300 N for a stem of 120 mm to 250 mm centre-to-tip length, 1 200 N above and below that band, with six specimens unbroken.

At a glance

Test type
Fatiguea load is applied over and over until something fails
Published by
ISO
Edition
ISO 7206-4:2010

From the test method to your testing system

Explore DAK equipment for ISO 7206-4, then review the specimen and setup requirements below.

Series 7200 Universal Testing MachineUniversal Testing MachineSeries 7200Explore the machine →Series 9000 Universal Testing MachineUniversal Testing MachineSeries 9000Explore the machine →
Jump to a specific section

01Understand the method

What the test does

The stemmed femoral component is embedded to a level the standard defines and loaded cyclically through the head at angles specified in two planes, reproducing the hip joint reaction during gait. The waveform is sinusoidal, the frequency between 1 Hz and 30 Hz. Forces and cycles are both specified: 5 000 000 cycles at an endurance limit force FD of 2 300 N for a stem of 120 mm to 250 mm centre-to-tip length, 1 200 N above and below that band, six specimens unbroken. Displacement amplitude is logged throughout, and the component either completes the cycles or the cycle count and failure location are recorded.

What it measures, and why it matters

Whether the stem survives a lifetime of walking. What sets the method apart from most implant fatigue testing is that the requirement is in the standard: endurance forces and cycle counts are specified rather than chosen by the applicant, which makes results comparable between devices. Geometry carries the rest of the meaning: the stem is a cantilever, so the embedded level fixes the lever arm and hence the bending moment, and the specified angles put the component into combined bending and torsion as a femur does.

02Prepare the specimen and test settings

The embedding level is the test

How much of the stem is held, and how much is free, decides the bending moment the implant sees. It is specified, not chosen.

Embedding level
Defined by the standard relative to the stemEmbedding deeper shortens the lever arm and reduces the moment. The result would look better and mean nothing.
Loading angles
Specified in two planesThey reproduce the orientation of the femur under gait loading rather than a convenient vertical.
Test forces
Specified, together with the number of cyclesUnusually, the requirements for the endurance limit are in the standard rather than left to the submission.
Simulates
The dynamic loading of a hip stem during gait
Part 6
Covers the neck region separately
Check the embedding medium has not degraded
Over a long runDakA medium that softens or cracks during millions of cycles changes the support and therefore the moment, mid-test.

This is a specified-force, specified-cycles endurance test. That makes it a pass or fail against a stated requirement rather than a characterisation, which is unusual among implant fatigue methods.

Test speed

Frequency
1 Hz to 30 Hz, sinusoidal at the primary frequencyClause 6.1 a) and 8.9. The band is a choice, not a fixed rate: run as fast as the fixture stiffness allows while still holding the prescribed load to ± 2 % of the load range. That choice sets the calendar — 5 000 000 cycles takes about 46 h at 30 Hz, just under 12 days at 5 Hz and roughly 8 weeks at 1 Hz.
Frequency — modular stems in fluid
5 Hz or lessClause 8.7. Modular components run submerged in fluid test medium held at 37 °C ± 1 °C and continuously aerated, and that protocol caps the frequency.
Endurance limit force FD and cycles
1 200 N for CT ≤ 120 mm; 2 300 N for 120 mm < CT ≤ 250 mm; 1 200 N for CT > 250 mm — all over 5 × 10⁶ cycles, with six specimens unbrokenTable 2, Clause 9, where CT is the stem centre-to-tip length. The 1 200 N figures are stated as lower limits that may need adjusting upwards on design, material and clinical grounds.
Minimum load in the cycle
200 N to 300 NThe note to 8.8 gives this as the range found necessary for satisfactory operation of the testing machine.
Load accuracy
± 2 % of maximum load, verified to ISO 4965
Reported
Survival to the specified cycles at the specified force, or the cycle at failure
Monitoring
Displacement amplitude through the runA rising amplitude usually means a crack has started well before the stem parts.
Deflection cut-off
Stop the test if deflection exceeds 1,25 × f or 5 mm, whichever is greaterClause 8.10. f is the initial vertical load axis deflection, and the threshold is set from a reading taken after about 300 cycles. Deflection is monitored to ± 0,2 mm.
Verify the applied force dynamically
At the test frequency and waveform actually usedDakA frame calibrated statically can under-deliver at frequency, and every one of five million cycles inherits the error.

03Build the test setup on a DAK machine

What the machine must be capable of

Sinusoidal cyclic loading anywhere from 1 Hz to 30 Hz, held within ± 2 % of the load range, applied load error no greater than ± 2 % of maximum load to ISO 4965. The frequency is chosen rather than fixed, and the choice sets the calendar: five million cycles takes about 46 hours at 30 Hz, under twelve days at 5 Hz, roughly eight weeks at 1 Hz. Aim at the top of the band for a dry room-temperature test on a non-modular stem. Modular stems are the exception — submerged in fluid at 37 °C ± 1 °C, continuously aerated, capped at 5 Hz or less.

Force must also be verified dynamically at the test frequency, not only under static calibration. ISO 7500-1 states that its values are not necessarily valid for dynamic testing, and a frame that under-delivers at frequency passes that error into every one of five million cycles. Displacement logging to ± 0,2 mm is needed too: it identifies crack initiation, and the machine must stop itself when deflection exceeds 1,25 times the initial value or 5 mm, whichever is greater — a threshold set from a reading taken after about 300 cycles.

Running ISO 7206-4 on the Series 7200 and Series 9000

Dak verifies against whichever standard the method names, and where a class applies our frames sit a class tighter than it asks.

The method asks forDak supplies
CapacityModerate — hip stem endurance forces are specified in the standard and run to a few kilonewtonsLoad cells from 1 kg to 60 ton on the Series 7200, and 0.5 to 100 kN on the Series 9000
Force accuracyDynamic force verification at the test frequencyISO 7500-1 Class 0.5, verified to ASTM E4, DIN 51221 and BS 1610
GrippingAn embedding medium holding the stem to a defined level, with the load applied at defined anglesOur a fixture built for this method, built to the specimen
Environment23 ± 2 °C standard laboratory atmosphere3009 series chambers, −150 °C to +400 °C — temperature only

04Run the test

How the test runs

  1. Identify the size and design of the stemmed femoral component.
  2. Prepare the embedding medium and fixture to the specified geometry.
  3. Embed the stem to the level the standard defines, checking it before the medium sets.
  4. Set the loading angles in both specified planes.
  5. Select a load cell suited to the specified test force.
  6. Verify the applied force dynamically at the test frequency.
  7. Apply the specified cyclic force.
  8. Log displacement amplitude continuously.
  9. Run to the specified number of cycles, or to failure.
  10. Record survival, or the cycle count and location of failure.
  11. Inspect the embedding medium afterwards for degradation.

05Calculate, report and interpret

Calculations

Bending moment on the stemM

M = F × lever arm, set by the embedding level and the load angles

This is why the geometry is specified so tightly. The force alone does not define the loading.

Endurance requirement

Survival at the specified force for the specified number of cycles

Both values come from the standard, which is what makes this a requirement rather than a characterisation.

Detecting a crack

A progressive increase in displacement amplitude at constant force

Worth logging continuously — a stem that cracks at two million cycles and parts at four has failed at two.

What the report has to contain

  • Reference to ISO 7206-4, the edition and whether Amendment 1:2016 applied
  • Component identification, size, material and surface finish
  • Embedding medium and the level achieved
  • Loading angles in both planes
  • Test force and frequency
  • How dynamic force was verified
  • Number of cycles completed, or the cycle at failure
  • Location of any fracture
  • Displacement amplitude history
  • Condition of the embedding medium after the run

What goes wrong in practice

Embedding to the wrong level, which flatters the result silently. Verifying force statically and running dynamically. Reporting only final fracture when the displacement trace shows the crack started far earlier. Running a non-metallic specimen near 30 Hz, where the standard warns frequency may itself influence the result, or a modular stem above the 5 Hz its fluid-immersed protocol allows. And omitting the loading angles from the report, without which the bending moment cannot be reconstructed from the force — a reviewer cannot check a result whose geometry is not stated, however complete the force data looks.

06Compare methods and find answers

ISO 7206-4 or ISO 7206-6

Part 4Part 6
RegionThe stem, embeddedThe neck region
QuestionDoes the stem survive gait loading?Does the neck survive it?
GeometryEmbedded to a defined levelSupported to expose the neck
Run togetherCommonlyCommonly

A stem and its neck fail in different places for different reasons, and a component can pass one and fail the other. Submissions normally carry both.

Questions we are asked about this test

What is ISO 7206-4?

It is the ISO endurance test for the stemmed femoral component of a hip prosthesis. The stem is embedded to a defined level, loaded cyclically through the head at specified angles to simulate the dynamic loading of gait, and required to survive a specified force for a specified number of cycles. The current edition is ISO 7206-4:2010 with Amendment 1:2016.

Why does the embedding level matter so much?

Because it sets the bending moment. The stem is a cantilever: the embedded portion is the support and the free portion is the lever arm, so embedding deeper shortens the arm and reduces the moment for the same applied force. A stem embedded a few millimetres too deep would survive a test it should fail, and nothing in the force record would show it. That is why the level is specified rather than left to the laboratory.

Why are the loading angles specified in two planes?

Because a femur is not loaded vertically. The hip joint reaction passes through the head at an angle in both the frontal and sagittal planes during gait, putting the stem into combined bending and torsion rather than simple bending. Reproducing those angles is what makes the test represent walking rather than a convenient laboratory arrangement.

Is this a characterisation or a pass-fail test?

A pass or fail. Unusually among implant fatigue methods, the standard specifies both the endurance limit forces and the corresponding cycle count: 5 000 000 cycles at 2 300 N for a stem of 120 mm to 250 mm centre-to-tip length and 1 200 N outside that band, with six specimens required unbroken. The requirement lives in the standard rather than in the submission, which makes results more directly comparable between devices than a method where the applicant chooses the loading.

Why monitor displacement rather than just wait for fracture?

Because a stem that has cracked has already failed, even if it has not parted. A fatigue crack reduces stiffness, so the displacement amplitude at constant force rises progressively before final fracture. Logging it continuously identifies the cycle at which the crack initiated, which is the meaningful number — a stem that cracks at two million cycles and breaks at four failed at two.

Why check the embedding medium afterwards?

Because it is part of the load path for the whole run. A medium that softens, creeps or cracks during several million cycles changes how the stem is supported, which changes the effective lever arm and therefore the moment — mid-test, and invisibly. Inspecting it afterwards is what confirms the loading was what the report claims it was.

Why verify the force dynamically?

Because a static calibration does not describe a machine in motion. ISO 7500-1 states that its values are not necessarily valid for high-speed or dynamic testing, and a frame that under-delivers at test frequency passes that error into every one of several million cycles. Nothing in the record afterwards reveals it, so the verification has to happen at the frequency and waveform actually used.

Materials tested to it

Industries that test to it

Other standards explained

Planning ISO 7206-4 testing?

Discuss your specimen, test requirements and reporting needs with DAK engineering.

Discuss your test setup →

This page describes the method as practised. The governing text is the current edition from the issuing body. Tell us what you are testing and we will answer with the machine, the fixture and a quotation.