What is ISO 11607-1?+
ISO 11607-1 specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems intended to keep terminally sterilised medical devices sterile until the point of use. The current edition is ISO 11607-1:2019, the second edition, amended by ISO 11607-1:2019/Amd 1:2023, Application of risk management, and adopted in the United States as ANSI/AAMI/ISO 11607-1:2019 with that amendment. Part 2 covers validation of the forming, sealing and assembly processes.
Is it a test method?+
No, and this is the most useful thing to know about it. It is a requirements standard: it says what a sterile barrier system must achieve and points to test methods that demonstrate it. Where the system is formed by sealing, it requires that specified requirements for seal width and seal strength are met, and seal strength is then measured by a referenced method — ASTM F88/F88M in most laboratories.
Why is there an upper limit on seal strength as well as a lower one?+
Because the pack has two jobs that conflict. It must maintain sterility until use, which argues for the strongest seal possible, and it must open cleanly at the bedside without tearing the web or shedding fibres into a sterile field, which argues for the weakest seal that will hold. Seal strength is where those two are reconciled, so a specification normally sets both limits and a value above the upper one is a failure just as a value below the lower one is.
Why does the failure mode matter?+
Because it tells you how the pack will open. An adhesive peel, where the two webs separate cleanly at the interface, is what a peelable pouch is designed to do. Delamination of the paper or fibre tear means the seal is stronger than the substrate: the pack opens by ripping rather than peeling and can shed particles at exactly the wrong moment. Two packs with the same peel force and different failure modes are not equivalent.
Where should specimens be taken from?+
From every edge of the seal. A pouch seal is formed by different parts of the sealing tooling on each side, and consistency between them is one of the things being validated. Cutting all the specimens from one convenient edge produces a tight, reassuring set of numbers about that edge, and the pack will fail at whichever edge nobody sampled.
Why is ageing part of the requirement?+
Because the shelf-life claim is. A sterile barrier system is labelled with an expiry date, and the evidence for that date has to include seal strength after real-time or accelerated ageing and after whatever sterilisation process the product goes through. Ethylene oxide, gamma and steam all change the materials to different degrees. The as-made figure says only that the pack was sealed correctly today.
What machine does it need?+
A small frame with a load cell sized in newtons. Seal peel forces are typically a few newtons and can be under one, so force accuracy to ISO 7500-1 Class 1 over that low range is the requirement and capacity is irrelevant. Grips must hold a thin film and a coated paper without cutting either, and the result is taken as an average over the peel length, so continuous logging and enough travel to complete the peel in one pass are both needed.